January 2009 UDA2182 Universal Dual Analyzer Product Manual 221
problem is that the cursor is positioned too far to the left or right of the flat portion of
the curve, move the cursor back to the flat portion of the curve.
15.
Perform another Air Calibration to correct any changes that occurred during the PBT.
16.
If you reached this point, you have both a working probe and analyzer that are
calibrated to one another correctly.