Honeywell EC-P2 Instruction and Installation Manual
8
System Specifications
Primary Power:
Alkaline D cell battery, 75 hours operation.
Note: An audible alarm single beep every 60 seconds
indicates that either the Alkaline D cell battery is low or
the internal rechargeable NiCad battery is low.
Standby Power:
Internal rechargeable NiCad battery, 6 hours
operation.
Note: An audible alarm single beep every 60 seconds
indicates that either the Alkaline C cell battery is low or
the internal rechargeable NiCad battery is low.
Charger:
120 VAC plug-in charger for internal standby battery
Display:
Backlit LCD
Pump:
Miniature internal diaphragm type, flowrate 0.5
SCFH
Alarms:
Two concentration alarms, zero drift caution, Low flow
and Low battery alarms. Alarms displayed on LCD and
indicated by audible beeper.
Standard Accessories:
Storage case, 10” extension wand, battery charger,
filter material, datalogging software, RS-232 output
cable (DB9 connector)
Optional Accessories:
Analog output cable. Sensor Keeper for storage of
additional SMART-CELLs
Sensor Type:
Target gas selective electrochemical SMART-CELL
Accuracy:
Sensor dependant but generally ± 5% of reading
Repeatability:
± 1% full scale
Operating Ambient Temperature
Range:
0° to +120° F
Storage Temperature:
Recommend storing on the charger in an air
conditioned office
Weight:
3 lbs. (shipping weight), 1.8 lbs. (detector only)
Dimensions:
Approx 3.5” x 5” x 8.5”
Outputs:
RS-232 output of stored gas values.
Analog output 0 to 1.0 VDC (requires optional
analog output cable)
Memory:
12,000 data points, storage interval between points
programmable at 1, 5, 10, or 15 minutes. Typical
capacity: 8 days at 1 minute storage interval
2.1 Operating the Sensor
The Honeywell EC-P2 is shipped ready for use. To power up the instrument, push
the PWR button. The power up sequence displays first the self check, the sensor
full scale value, the downscale caution level, warning concentration setpoint, alarm
concentration setpoint, output ranges, software version, then the normal operating
screen.
The unit will then display the instantaneous gas concentration and is ready for
use.