Honeywell SC1000 User Manual Page 107

  • Download
  • Add to my manuals
  • Print
  • Page
    / 174
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 106
SC Series Instruction Manual page 107
High-Level Input Channel12
12.4 Calibration Procedure
If you are not familiar with operating the instrument in the SETUP menu mode,
see “SETUP Menu mode” on page 23. A listing of all menu items is given in
“Setup Menu Reference” on page 157.
Step 1: Wire the transducer to the channel’s connector.
See “Wiring” on page 98 for details.
Step 2: Set the Excitation and Signal jumpers appropriate for the transducer,
amplifier or DC-DC LVDT.
See “Wiring” on page 98 and “Excitation and Signal Jumpers” on page 106.
Step 3: Enter the
CALIBRATION TYPE
.
There are two methods that can be used to calibrate the transducer to the
Input Channel. Each has advantages and disadvantages as described in
CALIBRATION TYPE
Menu Item” on page 113. It is important to know your
application in order to select the appropriate calibration type.
Step 4: Enter the
CALIBRATION DATA.
Otherwise, consult the Certificate of Calibration for the transducer when
entering information in the
CALIBRATION DATA
sub-menu.
Step 5: Perform the calibration.
Using the
CALIBRATE
menu item starts the calibration process. You will be
prompted to apply loads to the transducer as required.
Page view 106
1 2 ... 102 103 104 105 106 107 108 109 110 111 112 ... 173 174

Comments to this Manuals

No comments